• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE; SCS IPG Back to Search Results
Model Number 3660
Device Problems Wireless Communication Problem (3283); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/17/2019
Event Type  malfunction  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient underwent an unrelated surgical procedure.Patient did not set the ipg into surgery mode as recommended.Thereafter, the ipg failed to establish communication with external device.Recovery efforts were unsuccessful and the ipg was confirmed inoperable.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information received identified that patient underwent surgical intervention on (b)(6) 2020 wherein, the ipg was explanted and replaced.Effective therapy was restored post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key9263793
MDR Text Key164621268
Report Number1627487-2019-12122
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Remedial Action Other
Type of Report Initial,Followup,Followup,Followup
Report Date 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/05/2018
Device Model Number3660
Device Catalogue Number3660
Device Lot Number5649573
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight54
-
-