| Catalog Number |
321-25-38 |
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was beginning to prepare the glenoid for the baseplate.He began reaming as usual and immediately after starting to ream the pilot tip of the reamer broke off and was left behind in the glenoid.He stopped reaming and took time to retrieve the small pilot tip that was still inside the bone.It took him 5-10 minutes, but he eventually removed it.And he continued preparing the glenoid and the case was finished in the normal fashion.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) based on capa(b)(4), the broken device reported was likely the result of applying a bending moment to the reamer during use, which led to brittle fracture of the pilot tip feature.(h6) evaluation codes: 3165, 1069.
|
| |
|