• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH CONN O/O SD TOP NTCH 5.5X5.5 T; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH CONN O/O SD TOP NTCH 5.5X5.5 T; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 179771555
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 07/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Visual examination found that the tulip head threads on the connector were torn.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.With the information provided, a definitive root cause for the torn threads on the conn o/o sd top ntch 5.5 x 5.5 t cannot be determined.Noted damage suggests that inadvertently cross threading of the set screw occurred upon insertion into the tulip head of the connector.The root cause of the set screw¿s drive feature stripping could not be determined from the sample and the information provided.A potential root cause may be not fully inserting the associated driver into the set screw during tightening.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Dr (b)(6), expedium si set screws ¿ 186715000, open-closed connector - 179755155: repeatedly stripping.Procedure was completed by using multiple locking screws.No patient complications.Operation was completed as expected.15 minute delay to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONN O/O SD TOP NTCH 5.5X5.5 T
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9265578
Report Number1526439-2019-52351
Device Sequence Number419253
Product Code NKB
UDI-Device Identifier10705034077819
UDI-Public10705034077819
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number179771555
Device Lot NumberNW176122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/31/2019
Initial Report FDA Received Date11/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Patient Weight75
-
-