Information contained in this report was previously submitted through 410psr on 22/apr/2019.A review of the device history record has been completed.No deviations or non-conformances noted device evaluation: analysis of the returned device identified: deformation, biological tissue color yellow and weight to the spec.Cloudy was observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii, and "recall".This is a known potential adverse event addressed in the product labeling.
|