| Catalog Number |
510038BER |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/10/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during an endovascular aortic aneurysm repair procedure, a catheter tip detached within the patient.The physician had acquired retrograde femoral artery access.During catheter manipulations over a stiff guidewire, the catheter kinked and detached within the patient's ipsilateral common iliac artery.The physician was able to successfully remove the catheter tip from the patient with a vascular snare device.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.No definitive root cause could be determined however, it is likely that significant force was applied to the device during use.The device history record was reviewed and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were found.
|
| |
|