| Catalog Number |
PHA050602A |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records indicated the lot met pre-release specifications.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the patient being implanted with a gore® tigris® vascular stent in the popliteal artery.A 0.35 guidewire was reportedly used during the procedure.It was reported during deployment the device released half way and suddenly locked not allowing the device to fully deploy.Reportedly the physician stated it is unknown what caused the partial deployment.The device was able to be pulled back into the 8fr sheath and withdrawn from the patient.No additional adverse events were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Added a1: patient identifier.Corrected g5: combination product (yes).
|
| |
|
Additional Manufacturer Narrative
|
|
H1: corrected type of report (changed to malfunction).H6: corrected conclusion code.Engineering evaluation state, based on the available returned device, the following observations were made: the device was kinked at multiple locations along the catheter, including near the strain relief.The lock slider was in the unlocked position.The outer sheath was locked distally, over the proximal edge of the stent.Approximately 11mm of the distal end of the stent was uncovered.Upon disassembly of the handle, the inner shaft was kinked and folded at the location of the pulley within the handle.Based on the device examination performed, no manufacturing deficiencies were identified.
|
| |
|
Search Alerts/Recalls
|