|
(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional evaluation summary: representative samples were received for product code 8533, lot kdh036.During the visual inspection of the samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were observed.A functional test was performed and the tensile strength force was above the minimum requirement.Per the condition of the representative samples, no suture breakage was found and the tested samples met the finished goods requirements.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What suture was used to complete the anastomosis? product code and lot number? what is the physician¿s opinion as to the etiology of or contributing factors to the patient¿s death? the patient demographic info: age, gender, weight, bmi at the time of index procedure date of the index surgical procedure the diagnosis and indication for the index surgical procedure in what tissue was the suture placed? what was the tissue condition, i.E., normal or thin, calcified, fragile, diseased? what suture was used to complete the anastomosis? product code and lot number? date of patient death? was an autopsy performed? if yes, can you provide report of findings? other relevant patient history/concomitant medications, lot number.Device return status?.
|