WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 9MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 02.130.209
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional procode: hrs.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown surgery on (b)(6) 2019, one of the locking holes of the variable angle (va) locking t-plate was not accepting a va locking screw and just completely bent.The surgeon tried another va locking screw and still it would not lock.However, the plate was implanted.It is unknown if there was a surgical delay.Procedure and patient outcome are unknown.This complaint involves three (3) devices.This is 3 of 3 for report (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device not returned for investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Concomitant device reported: unknown screw (part# unknown; lot# unknown; quantity# 1).
 
Event or Problem Description
It was clarified that the locking screw is what was bent.The plate was not bent.It was noted there was a two (2) minute surgical delay.The surgeon left one locking screw in despite not locking into the plate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 9MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9268185
Report Number2939274-2019-61897
Device Sequence Number11417005
Product Code HWC
UDI-Device Identifier10886982044631
UDI-Public(01)10886982044631
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K141527
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.130.209
Device Catalogue Number02.130.209
Was Device Available for Evaluation? No
Date Returned to Manufacturer11/13/2019
Initial Date Received by Manufacturer 10/16/2019
Supplement Date Received by Manufacturer10/28/2019
12/02/2019
Initial Report FDA Received Date11/01/2019
Supplement Report FDA Received Date11/15/2019
12/12/2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: TRAUMA
Patient Age79 YR
-
-