(b)(4).Date sent to the fda: 01/29/2020.A manufacturing record evaluation was performed for the finished device lot al8307, and no non-conformances were identified.Additional h3 investigation summary: it was reported breakage suture.One needle-suture piece of product code k834, lot al8307 was received for analysis.During the visual inspection of the opened sample, the swage and attachment area were noted to be as expected.However, marks on the body needle and the end of the suture piece cut that appears to be by the use of a surgical instrument could be observed.The other section of the suture was not returned for analysis.Additionally, a functional test was performed and the tensile strength force was above the minimum requirement.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling.Additional information was requested and the following was obtained: please provide the specific number of sutures that broke during this procedure.Customer confirmed there were 5 broken threads.Note: events reported in 2210968-2019-89406, 2210968-2019-89410, 2210968-2019-89414, 2210968-2019-89415 and 2210968-2020-00788.
|