| Model Number |
3CX*RX25RW |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
|
| |
|
Event or Problem Description
|
|
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the arterial thermistor did not record the temperature.As per user facility, the cable was fine when they tested it on the venous thermistor port and was able to record the temperature.No consequences or impact to patient.The product was not changed out.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 24, 2019.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).H4 (device manufacture date).H6 (identification of evaluation codes 11,3331,4114, 213, 67).Method code #1: 11 - testing of device from same lot/batch retained by manufacturer.Method code #2: 3331 - analysis of production records.Method code #3: 4114 - device not returned.Results code: 3259 - improper physical structure.Conclusions code #1: 4307 - cause traced to component failure.The affected sample not was returned for evaluation so a thorough investigation could not be conducted.A representative retention sample from the same product/lot number combination was tested and found to be functioning properly.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|
Search Alerts/Recalls
|