| Catalog Number |
1550350-28 |
| Medical Device Problem Codes |
Material Rupture (1546); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Vessel Or Plaque, Device Embedded In (1204)
|
| Date of Event |
10/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties and subsequent treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was performed to treat a heavily calcified lesion in the heavily tortuous right coronary artery.The 3.5x28 mm xience sierra stent delivery system (sds) was advanced, however the balloon ruptured during the first inflation to eight atmospheres.The stent dislodged and was successfully crushed to the vessel wall with a non-abbott stent.There was no adverse patient sequela or a clinically significant delay in the procedure.
|
| |
|