| Catalog Number |
1550350-18 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Vessel Or Plaque, Device Embedded In (1204)
|
| Date of Event |
10/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties and subsequent treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion located in the left anterior descending coronary artery.The xience sierra stent delivery system was advance in the patient, but failed to cross due to the heavily calcified anatomy.During removal, there was resistance with the anatomy, and the stent dislodged from the balloon.A non-abbott stent was deployed and crushed the stent to the vessel wall to complete the procedure.There was no adverse patient sequela or a clinically significant delay in procedure.No additional information was provided.
|
| |
|