• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY Back to Search Results
Model Number PNML6F088904M
Medical Device Problem Codes Leak/Splash (1354); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a medical procedure in the right middle cerebral artery (mca) using a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician found that the hemostasis valve adaptor (hva) was leaking and was not tightening properly; therefore, it was removed.The procedure was completed using another neuron max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the returned neuron max was not damaged.The hva was not damaged.Conclusions: evaluation of the returned neuron max and its hva revealed undamaged devices.During functional testing, the returned hva was flushed through with water and no leaking was observed.The reported complaint was unable to be confirmed.Evaluation of the returned neuron max and its hva revealed undamaged devices.During functional testing, the returned hva was flushed through with water and no leaking was observed.The reported complaint was unable to be confirmed.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9270629
Report Number3005168196-2019-02041
Device Sequence Number16677013
Product Code DQY
UDI-Device Identifier00814548012162
UDI-Public00814548012162
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/14/2022
Device Model NumberPNML6F088904M
Device Catalogue NumberPNML6F088904M
Device Lot NumberF89581
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2019
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer02/18/2020
Initial Report FDA Received Date11/03/2019
Supplement Report FDA Received Date03/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Date Report Sent to FDA01/01/2005
-
-