| Model Number |
PNML6F088904M |
| Medical Device Problem Codes |
Leak/Splash (1354); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a medical procedure in the right middle cerebral artery (mca) using a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician found that the hemostasis valve adaptor (hva) was leaking and was not tightening properly; therefore, it was removed.The procedure was completed using another neuron max.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the returned neuron max was not damaged.The hva was not damaged.Conclusions: evaluation of the returned neuron max and its hva revealed undamaged devices.During functional testing, the returned hva was flushed through with water and no leaking was observed.The reported complaint was unable to be confirmed.Evaluation of the returned neuron max and its hva revealed undamaged devices.During functional testing, the returned hva was flushed through with water and no leaking was observed.The reported complaint was unable to be confirmed.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.
|
| |
|