| Model Number |
8888145048CP |
| Medical Device Problem Codes |
Product Quality Problem (1506); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
06/17/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the patient was hospitalized due to tunnel infection, which was treated with vancomycin.It was also noted that the catheter was not repaired, there was no leak, tego was not utilized, and there was no luer adapter issue.A new catheter was inserted into the patient without silver coating, and the patient was eventually discharged and stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b2, b5, d6b, d8, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the first photo depicts the catheter inserted into a patient.The nodules on the silver sleeve are a normal part of the function of the device when the silver has been dissolved.The second photo depicts the catheter inserted into a patient.The nodules on the silver sleeve are a normal part of the function of the device when the silver has been dissolved.Without the physical product functional evaluation was unable to be performed.It was reported that the catheter's coating was blistering.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the patient was hospitalized due to tunnel infection which was treated with intravenous (iv) vancomycin.It was stated that air bubbles were still noted on the silver coating of the device.It was also noted that the catheter was not repaired, there was no leak, tego was not utilized, and there was no luer adapter issue.A new catheter was inserted to the patient without silver coating, and the patient was eventually discharged and stable.
|
| |
|