MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
The complained test strips have been calibrated against the who standard and are in scope of the roche initiated recall.For these test strips, there is a potential for a product problem when the inr is > 4.5.Values > 4.5 inr showed an increasing positive bias.The information in the case is consistent with the details of the recall and the issue has been fully investigated.Per product labeling, coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.Occupation ni equals lay user / patient.
 
Event Description
The initial reporter complained of a questionable high inr result from a coaguchek xs meter serial number (b)(4) compared to an unknown laboratory method.The meter result was 5.3 inr and within 4 hours the laboratory result was 3.3 inr.The patient's therapeutic range is 2.0 - 3.0 inr.The laboratory result was deemed to be correct.The patient had a recent decrease in their dosage.The patient had a hematocrit of 30.7 percent.The patient does take plavix.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key9270730
MDR Text Key219591303
Report Number1823260-2019-03934
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2020
Device Catalogue Number04625374160
Device Lot Number30497423
Was Device Available for Evaluation? No
Date Manufacturer Received10/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberRES Z-0360-2019
Patient Sequence Number1
Patient Age65 YR
Patient Weight91
-
-