| Model Number |
21AECJ-502 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2019, a 21 mm masters valve was selected for implant.While rotating the valve, the leaflet fractured.The device was explanted and replaced with a 19 mm masters valve.The patient remained hemodynamically stable but there was a clinically significant delay in the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of the leaflets fractured during rotation was confirmed.One leaflet had fractured into multiple pieces.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the fractured leaflet.The cause of the reported event could not be conclusively determined, however it is consistent with damage caused by some external force applied to the leaflet and orifice, which overstressed the carbon material.
|
| |
|