Model Number 21AECJ-502
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/19/2019
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2019, a 21 mm masters valve was selected for implant.While rotating the valve, the leaflet fractured.The device was explanted and replaced with a 19 mm masters valve.The patient remained hemodynamically stable but there was a clinically significant delay in the procedure.
 
Additional Manufacturer Narrative
The reported event of the leaflets fractured during rotation was confirmed.One leaflet had fractured into multiple pieces.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the fractured leaflet.The cause of the reported event could not be conclusively determined, however it is consistent with damage caused by some external force applied to the leaflet and orifice, which overstressed the carbon material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key9271281
Report Number2648612-2019-00091
Device Sequence Number1892864
Product Code LWQ
UDI-Device Identifier05414734053884
UDI-Public05414734053884
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/01/2024
Device Model Number21AECJ-502
Device Catalogue Number21AECJ-502
Device Lot Number7075208
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/01/2019
Supplement Date Received by Manufacturer12/06/2019
Initial Report FDA Received Date11/04/2019
Supplement Report FDA Received Date12/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient Weight75
-
-