Model Number 3660 |
Device Problem
Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 10/16/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported the patient was unable to establish a connection between the ipg and the patient controller (pc).The patient stated they underwent a surgical procedure and did not place the device into surgery mode.A company representative met with the patient and confirmed the issue by also using the clinician programmer (cp).The representative attempted multiple times to connect to the ipg, but to no avail.In turn, the patient may undergo surgical intervention as the next course of action.
|
|
Manufacturer Narrative
|
The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery was used.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.The device was not returned for analysis; however, the implant data log was received.Analysis of the record shows that the system was unable to successfully maintain a connection with the clinician programmer.Actions have been taken to prevent reoccurrence.
|
|