| Model Number |
FSC330-G7529 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A fuse scope was received for analysis.A visual analysis was performed and found an internal wiring failure of the charge-coupled device (ccd) circuit at the distal tip.The ccd was replaced to resolve the issue.The reported issue was confirmed when the scope was angulated at the distal tip.Based on a thorough review of the reported complaint, the most probable cause for this complaint is cause traced to component failure.This complaint investigation conclusion code was utilized for an expected or random component failure without any design or manufacturing issue.Based on the results of this investigation, no escalation is necessary.The repair history was reviewed and nothing was found to indicate a possible service-related cause for the complaint.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a fuse c38s slim colonoscope - s was used during a colonoscopy procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the images turned black when the scope was angulated.The anatomy of the patient was not visible on center image when the issue occurred.Another fuse c38s slim colonoscope - s was used to complete the procedure.There were no patient complications reported as a result of this event.
|
| |
|