• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SOLERA SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SOLERA SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 6534530
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent posterior fixation at l1-l3 due to l2 chance fracture.Intra op, the set screw could not be inserted and its threads got damaged on insertion at left side of l3.The set screw was then explanted.It is said that the screw at that location was angled compared to the other screws.There was a delay of less than 60 minutes in overall procedure time as a result of this event.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual and macroscopic examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SOLERA SPINAL SYSTEM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key9272721
Report Number1030489-2019-01263
Device Sequence Number11236007
Product Code NKB
UDI-Device Identifier20643169885831
UDI-Public20643169885831
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K170679
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6534530
Device Lot NumberH5544544
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Initial Date Received by Manufacturer 10/11/2019
Supplement Date Received by Manufacturer04/13/2020
Initial Report FDA Received Date11/04/2019
Supplement Report FDA Received Date05/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR
Patient Weight80
-
-