• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; DXE Back to Search Results
Model Number PMX110
Medical Device Problem Code Decrease in Suction (1146)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the popliteal vein using a penumbra system aspiration pump max 110 (pump max).During the procedure, it was reported that the flow through the indigo system aspiration catheter 8 (cat8) and into the indigo pump max canister (canister) was very slow.The physician then found that the pump max was unable to produce maximum aspiration and was only able to reach -18 in hg.Therefore, the physician decided to remove the pump max from the procedure.The procedure was completed using a new pump max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: a rattling noise was observed while handling the pump max and, upon inspection of the bottom of the pump, one of the screw ports was observed to be damaged and the screw was missing.Conclusions: evaluation of the returned pump max revealed a functional device.During functional testing, the pump max was powered on and was able to achieve vacuum pressure within specification.The reported complaint could not be confirmed.Further evaluation revealed a rattling noise within the pump max that was determined to be a missing screw from the bottom of the pump.This damage was likely incidental to the reported complaint and may have occurred during an attempt to open the pump¿s housing.No other devices associated with the complaint were returned for evaluation.Penumbra pumps are 100% visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9273207
Report Number3005168196-2019-02044
Device Sequence Number18872920
Product Code DXE
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (Y/N)Y
Initial Reporter CountryCO
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/07/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMX110
Device Catalogue NumberPMX110
Device Lot NumberF06970-36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/07/2019
Supplement Date Received by Manufacturer01/30/2020
Initial Report FDA Received Date11/04/2019
Supplement Report FDA Received Date01/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Date Report Sent to FDA01/01/2005
-
-