U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE S MRI SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE S MRI SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 693552
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the lead warning triggered, and the right ventricular (rv) lead exhibited a sudden increase in rv coil impedances.The lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular defibrillation coil was beyond the expected upper range.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key9273412
Report Number2649622-2019-20224
Device Sequence Number19441292
Product Code LWS
UDI-Device Identifier00613994446336
UDI-Public00613994446336
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/01/2012
Device Model Number693552
Device Catalogue Number693552
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2019
Supplement Date Received by Manufacturer11/04/2019
Initial Report FDA Received Date11/04/2019
Supplement Report FDA Received Date11/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DVFB1D1 ICD
Patient Age19 YR
-
-