COOK IRELAND LTD ZILVER PTX 35 DRUG-ELUTING STENT Back to Search Results
Model Number G38463
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Common name and pro-code: unknown.
 
Event or Problem Description
According to the initial reporter, patient was in interventional radiology for ir thrombolysis therapy and angiogram.During the procedure a zilver ptx 6mmx 40mm stent was used and it prematurely deployed.While this stent was bring retracted, using the filter and the filter wire, part of the stent sheared off near the left common femoral artery access site.Ct was ordered to determine exact location of sheared off site.Per ct: no definite intravascular foreign body is visualized.Questionable hyperdense fractured stent material surrounding the sheath.Patient was subsequently transferred to (b)(6) (account (b)(6) ) in the event further intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER PTX 35 DRUG-ELUTING STENT
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9273979
Report Number3005580113-2019-00717
Device Sequence Number181627
Product Code NIU
UDI-Device Identifier10827002384634
UDI-Public(01)10827002384634(17)210115(10)C1613840
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/04/2019
Report Date (Section F) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2021
Device Model NumberG38463
Device Catalogue NumberZISV6-35-125-6-40-PTX
Device Lot NumberC1613840
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/29/2019
Event Location Hospital
Date Report to Manufacturer11/04/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA11/04/2019
-
-