• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUNE CRS FB TIB BASE SZ 7 CEM; ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUNE CRS FB TIB BASE SZ 7 CEM; ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150640007
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address laxity and loosening of the tibial tray at the cement to implant interface.Depuy cement was used.It was also reported that the laxity is present in both flexion and extension.The surgeon planned to put in a thicker insert.The surgeon decided also to revised the femur.The patella was retained and was well fixed.The femur was well fixed, but the patient needed better extension, so revising the femur and moving it proximal was the best plan.No cement stuck to the underside of the tibial tray.Doi: (b)(6) 2018.Dor: (b)(6) 2019.Left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUNE CRS FB TIB BASE SZ 7 CEM
Common Device Name
ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9274756
Report Number1818910-2019-113964
Device Sequence Number19817015
Product Code JWH
UDI-Device Identifier10603295042815
UDI-Public10603295042815
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K160700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number150640007
Device Lot Number8589175
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2019
Supplement Date Received by Manufacturer12/04/2019
Initial Report FDA Received Date11/04/2019
Supplement Report FDA Received Date12/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-