| Model Number |
CD3257-40Q |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/15/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
| |
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient condition was stable before, during, and after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Premature depletion was not confirmed by analysis.However, an abnormal transient battery voltage drop was detected.The voltage drop is consistent with li deposit in the battery.
|
| |
|