• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3257-40Q
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
 
Event or Problem Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient condition was stable before, during, and after the procedure.
 
Additional Manufacturer Narrative
Premature depletion was not confirmed by analysis.However, an abnormal transient battery voltage drop was detected.The voltage drop is consistent with li deposit in the battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key9276512
Report Number2938836-2019-15754
Device Sequence Number19430370
Product Code NIK
UDI-Device Identifier05414734505635
UDI-Public05414734505635
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2014
Device Model NumberCD3257-40Q
Device Catalogue NumberCD3257-40Q
Device Lot Number4006541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/16/2019
Supplement Date Received by Manufacturer01/21/2020
Initial Report FDA Received Date11/04/2019
Supplement Report FDA Received Date01/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0004-2018;Z-0116-2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-