|
Model Number 10013361T |
Device Problems
Stretched (1601); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: 1000ml baxter bag lot y300392 exp sep 20, doxorubicin (adriamucin) 16.4mg, etoposide (toposar) 82mg in sodium chloride 0.9%, 1000ml chemo infusion;10ml zr swabflush syringe lot 3136603 exp 2020-05-01 0.9% nacl injection; td 10/05/2019 the devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
|
Event Description
|
It was reported that the tubing set developed a "bubble" in the silicone segment while attached to a combined chemotherapy infusion bag of doxorubicin (adriamucin) 16.4m, etoposide (toposar) 82mg/sodium chloride 0.9%, 1000ml to infuse at a rate of 44.3ml/hour for a duration of 24 hours with a total vtbi of 1062.3ml.The customer further stated that there were no iv push medications administered prior to the balloon developing in the tubing set.The customer later stated that there were no adverse effects caused to the adult patient from the event.
|
|
Event Description
|
It was reported that the tubing set developed a "bubble" in the silicone segment while attached to a combined chemotherapy infusion bag of doxorubicin (adriamucin) 16.4m, etoposide (toposar) 82mg/sodium chloride 0.9%, 1000ml, to infuse at a rate of 44.3ml/hour for a duration of (24) hours, with a total vtbi of 1062.3ml.The customer further stated that there were no iv push medications administered prior to the "balloon" developing in the tubing set.The customer later stated that there were no adverse effects caused to the adult patient from this event.
|
|
Manufacturer Narrative
|
The customer¿s report of a bulge in the silicone segment was confirmed by visual inspection of the as-received sample.Since the observation of the balloon in the silicone tubing segment has already been observed in a high number of previously investigated complaints, functional testing was not conducted.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection confirmed a balloon in the silicone pump segment near the upper fitment.The silicone pump segment was found to be within measurement specification(s) and the tubing was visually confirmed to be concentric.Previously investigated complaints for this failure mode determined that the ballooning is caused by excess pressure within the silicone tubing segment.The root cause for the source of the excessive pressure is unknown.
|
|
Search Alerts/Recalls
|
|
|