• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6020-0230
Medical Device Problem Codes Degraded (1153); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 10/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
As reported: "[left hip] revision surgery.Revised head ball, liner, dome hole plug, screw and stem.Acetabular cup was left and inspected - deemed intact.New liner and dome hole plug implanted.Head was disassociated off trunnion.Stem was unusually worn down.Stem revised and explanted as well as head.Unusual wear on stem trunnion".Rep provided pre- and post-revision x-rays, an excerpt of the operative report, and pictures of the explants.The liner pictured shows a noticeably damaged rim in one spot.Spoke to rep.Liner was in that condition in vivo.The devices are allegedly available for return, but no further information will be available.
 
Additional Manufacturer Narrative
An event regarding disassociation and wear involving an accolade stem was reported.The event was confirmed by visual analysis of the devices.Method & results: device evaluation and results: damage consistent with the loss of taper lock was observed on the stem neck and trunnion.No identifiable materials or manufacturing discrepancies were observed on the surfaces examined for the accolade stem.Clinician review: a review of the provided x-rays by a clinical consultant indicated: based upon the information available for review, no determination can be made regarding the cause of this clinical event occurring ten years post- implantation.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports and progress notes are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
As reported: "[left hip] revision surgery.Revised head ball, liner, dome hole plug, screw and stem.Acetabular cup was left and inspected - deemed intact.New liner and dome hole plug implanted.Head was disassociated off trunnion.Stem was unusually worn down.Stem revised and explanted as well as head.Unusual wear on stem trunnion".Rep provided pre- and post-revision x-rays, an excerpt of the operative report, and pictures of the explants.The liner pictured shows a noticeably damaged rim in one spot.Spoke to rep.Liner was in that condition in vivo.The devices are allegedly available for return, but no further information will be available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE TMZF HIP STEM #2
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9278164
Report Number0002249697-2019-03678
Device Sequence Number18918293
Product Code LPH
UDI-Device Identifier04546540510549
UDI-Public04546540510549
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2014
Device Model Number6020-0230
Device Catalogue Number6020-0230
Device Lot Number30728201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2019
Initial Date Received by Manufacturer 10/09/2019
Supplement Date Received by Manufacturer12/30/2019
Initial Report FDA Received Date11/05/2019
Supplement Report FDA Received Date01/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age52 YR
-
-