MAUDE Adverse Event Report: COOK ENDOSCOPY ACUSNARE POLYPECTOMY SNARE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ACUSNARE POLYPECTOMY SNARE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number G22631
Device Problem Failure to Conduct (1114)
Patient Problem No Patient Involvement (2645)
Event Date 10/09/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: our laboratory evaluation of the product said to be involved could not confirm the report as it was described.The handle was manipulated and the snare head opened and close as intended.The handle of the device was manipulated and the snare head opened and closed with no resistance felt.During a visual inspection of the device, there were some slight bends throughout the length of the catheter.The continuity from the electrical pin to the snare head was tested with an ohm meter and passed.An additional functional test was performed by attaching the active cord to the electrical pin.The active cord connected to the device easily and remained securely connected.The device was connected to a valley lab generator and power was applied.The snare cut simulated tissue as expected.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for the reported observation could not be determined because the product said to be involved functioned as intended.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.The instructions for use contain the following information to assist with proper set-up and use of the device: "before using this device, follow recommendations provided by electrosurgical unit manufacturer to ensure patient safety through proper placement and utilization of patient return electrode.Ensure a proper path from patient return electrode to electrosurgical unit is maintained throughout the procedure." "inspect active cord.Cord must be free of kinks, bends, breaks and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." "with electrosurgical unit off, prepare equipment.Securely connect active cord to device handle and electrosurgical unit.Active cord fittings should fit snugly into both device handle and electrosurgical unit.Following instructions from electrosurgical unit manufacturer, position patient return electrode and connect it to electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify desired settings and activate electrosurgical unit.Note: maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During a colon polypectomy procedure, the nurse attempted to use a cook acusnare polypectomy snare to eliminate polyps.The wire didn't work at electrocautery function [unable to conduct current].The nurse used another lot of the same device and finished the procedure.On 04-nov-2019 clarification information was received noting the issue was found prior to insertion down the endoscope.This occurred prior to patient contact; there was no impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSNARE POLYPECTOMY SNARE
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key9278915
MDR Text Key219780510
Report Number1037905-2019-00670
Device Sequence Number1
Product Code KNS
UDI-Device Identifier00827002226319
UDI-Public(01)00827002226319(17)220513(10)W4214228
Combination Product (y/n)N
PMA/PMN Number
K851958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/13/2022
Device Model NumberG22631
Device Catalogue NumberASM-1-S
Device Lot NumberW4214228
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/10/2019
Initial Date FDA Received11/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-