| Model Number |
S208 |
| Medical Device Problem Codes |
Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
|
| Date of Event |
07/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker device was at end of life (eol) status.It was noted that high outputs were decreasing interval from elective replacement indicator (eri) to eol.This pacemaker device was explanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being filed to correct event problem and evaluation codes.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker device was at end of life (eol) status.It was noted that high outputs were decreasing interval from elective replacement indicator (eri) to eol.This pacemaker device was explanted.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the battery of this pacemaker was depleting prematurely.The device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being filed to correct event problem and evaluation codes.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.The returned pacemaker was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
| |
|