• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ALTRUA 20; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ALTRUA 20; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number S208
Medical Device Problem Codes Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
Date of Event 07/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
It was reported that this pacemaker device was at end of life (eol) status.It was noted that high outputs were decreasing interval from elective replacement indicator (eri) to eol.This pacemaker device was explanted.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This supplemental report is being filed to correct event problem and evaluation codes.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
It was reported that this pacemaker device was at end of life (eol) status.It was noted that high outputs were decreasing interval from elective replacement indicator (eri) to eol.This pacemaker device was explanted.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that the battery of this pacemaker was depleting prematurely.The device was explanted and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This supplemental report is being filed to correct event problem and evaluation codes.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.The returned pacemaker was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 20
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
MDR Report Key9279356
Report Number2124215-2019-24007
Device Sequence Number2483994
Product Code LWP
UDI-Device Identifier00802526484421
UDI-Public00802526484421
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S106
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/22/2011
Device Model NumberS208
Device Catalogue NumberS208
Device Lot Number592132
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/13/2019
Supplement Date Received by Manufacturer08/13/2019
10/31/2022
Initial Report FDA Received Date11/05/2019
Supplement Report FDA Received Date12/12/2019
11/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-