| Model Number |
H420995 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
During an ablation procedure, the 3d navigation did not function as expected.Troubleshooting was attempted by repositioning patches, revalidating, and restarting the computer and amplifier, with no resolution.Exchaning the navlink resolved the issue and the procedure was continued with no consequences to the patient.The length of procedure was extended by an hour and was clinically significant due to increased heparin administration and difficulties in managing sedation for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One ensite velocity¿ navlink module was received for investigation.Visual inspection of the returned product confirmed all input and output connectors were free of physical damage.Functional testing performed verified circuit integrity on all channels during the evaluation period.Based on the information provided to abbott and the investigation performed, the root cause of the reported display issue and subsequent delay in procedure cannot be conclusively determined as no abnormalities were identified.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|
| |
|