MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7134
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event of (b)(6) 2019 is an estimated date based off the aware date as the exact event date was not reported.
 
Event or Problem Description
It was reported that the balloon had deflation issue.A 2.50mm x 20mm emerge balloon was selected for use.At the end of the procedure, it was noted that the attachments to the indeflator seemed to slip off and not stay attached.Subsequently, the emerge balloon was not deflating easily.The device was successfully removed from the patient's body.No patient complications were reported.
 
Event or Problem Description
It was reported that the balloon had deflation issue.A 2.50mm x 20mm emerge balloon was selected for use.At the end of the procedure, it was noted that the attachments to the indeflator seemed to slip off and not stay attached.Subsequently, the emerge balloon was not deflating easily.The device was successfully removed from the patient's body.No patient complications were reported.
 
Additional Manufacturer Narrative
Date of event of (b)(6) 2019 is an estimated date based off the aware date as the exact event date was not reported.Device evaluated by mfr: the device was returned for evaluation.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was soaked in a water bath for 18 days to loosen the blood and contrast in the device.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.The device was able to inflate and deflate normally.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9280621
Report Number2134265-2019-13290
Device Sequence Number14014389
Product Code LOX
UDI-Device Identifier08714729806325
UDI-Public08714729806325
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7134
Device Catalogue Number7134
Device Lot Number0024157592
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2019
Initial Date Received by Manufacturer 10/11/2019
Supplement Date Received by Manufacturer12/09/2019
Initial Report FDA Received Date11/05/2019
Supplement Report FDA Received Date12/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-