| Model Number |
MMT-723NAL |
| Medical Device Problem Codes |
Mechanical Problem (1384); Power Problem (3010); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.The customer¿s blood glucose level was unknown.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with motor error alarmed during rewind due to motor encoder signal out of phase.Unable to perform operating currents, self-test, a21 error test, rewind test, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and displacement test or verify failed battery alarm due to motor error alarms.
|
| |
|