ABBOTT POINT OF CARE I-STAT TROPONIN (CTNI) CARTRIDGE; CTNI CARTRIDGE Back to Search Results
Catalog Number 06P23-25
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Chest Pain (1776)
Date of Event 10/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin result of 0.00 on a (b)(6) year old female patient with chest pain.There was no additional patient information at the time of this report.Return product is available for investigation.Method: i-stat, date: 10/28/2019, collected: 09:37, tested: 09:44, result: 2.51 ug/l, sample: a.I-stat, 10/28/2019, 15:40, 15:55, 0.00 ug/l, b.I-stat, 10/28/2019, 15:40, 16:07, 3.24 ug/l, b.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.There is evidence the patient suffered an mi.The investigation is underway.Per i-stat system manual: art: 714258-00q.Reportable range: 0.00 - 50.00.Reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results).Reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
 
Additional Manufacturer Narrative
Apoc incident # (b)(4).The investigation was completed on 12/16/2019.A review of the device history record (dhr) confirmed that the cartridge lot met finished goods (fg) release criteria.Returned cartridge testing met the acceptance outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).A deficiency for ctni cartridge lot g19198 has been identified.Retained cartridge test event ec155 met the acceptance criterion for the undetected error (udt) rate, but not for the detected error (dt) rate, as outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).Quality record (qr) 708510 will document the root cause investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT TROPONIN (CTNI) CARTRIDGE
Common Device Name
CTNI CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key9283904
Report Number2245578-2019-00260
Device Sequence Number2003204
Product Code MMI
UDI-Device Identifier10054749001986
UDI-Public10054749001986
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K031739
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/15/2020
Device Catalogue Number06P23-25
Device Lot NumberG19198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2019
Initial Date Received by Manufacturer 10/28/2019
Supplement Date Received by Manufacturer12/16/2019
Initial Report FDA Received Date11/06/2019
Supplement Report FDA Received Date12/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-