|
ENCORE MEDICAL L.P. X-ALT HIP; LINER, ACETABULAR, HOODED, NEUTRAL, 10°, HXL/FMP, 36MM ID, MP7
|
Back to Search Results |
|
| Model Number |
932-36-252 |
| Health Effect - Clinical Codes |
Failure of Implant (1924); Joint Dislocation (2374)
|
| Date of Event |
10/14/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred on same day.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are many factors that may contribute to the event that are outside the control of djo surgical.Due to the short time between previous and revision surgery, it is also possible that the event may have been occurred due to lack of post-operative care, patient noncompliance with medical instructions, improper surgical technique or incorrect implant selection.There are no indications of a product or process issue affecting implant safety or effectiveness.
|
| |
|
Event or Problem Description
|
|
Revision surgery - due to dislocation.
|
| |
|
Search Alerts/Recalls
|
|
|