(b)(4).A manufacturing record evaluation was performed for the finished device lot number paj128, and no non-conformances related to the reported complaint condition were identified.Investigation summary: unopened samples of product code m8705, lot # paj128 were returned for analysis.During the visual inspection of unopened samples, no defects were found on the packages.The samples were opened, and each packet contains four strands double armed.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the sample received, no pull off suture needle was found, and the tested sample met the finished goods requirements.
|