| Model Number |
TRF-240 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/15/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade iii and removal "due to the patient¿s concern with the product¿.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes. device evaluation: analysis of the returned device identified overweight device, crease fold and an opening on the anterior side.A visual and micro analysis was performed which identified: puncture opening.Based on the device analysis, the final assessment is a striated puncture due to surgical damage-like, consistent with the use of a surgical tool.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|