|
Product complaint # (b)(4).A manufacturing record evaluation was performed for the finished device batch mmh267 and no non-conformances were identified.The actual device was received for analysis a labeled winding former with a detached needle of product code 8631, mmh267.During the visual inspection of a detached needle, the swage and attachment area were noted to be as expected.Body fluids was present on needle and marks by use of a surgical instrument could be observed at the edge of the barrel hole, this condition causes the needle pull off.The suture was not received for evaluation.To avoid this kind of damage: grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the sample received, the assignable cause of the performance pull off - suture needle is an improper handling.Representative samples of an opened box with ten unopened samples of product code 8631, mmh267.During the visual inspection of ten unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the sample received, no performance pull off suture needle was found, and the tested sample met the finished goods requirements.
|