| Model Number |
367019 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pantera leo balloon catheter was chosen for treading a calcified lesion (90 percent stenosis degree) in the rca.After successful treatment, during withdrawal the device fractured.The distal part could be removed with a snare.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The complaint instrument was returned in two fragments.The hypotube has fractured about 23.3cm distal to the kink protector.Microscopic analysis of the fracture sites showed the cross-section is no longer circular which indicates that the hypotube most likely fractured as a result of significant bending.Other than reported the balloon of the complaint instrument has not been inflated.Review of the production documentations for the product detailed above verified that the instrument was manufactured according to specification and fulfilled all the requirements of in-process and final inspection.During final inspection and inner packaging every device is being visually inspected for kinks.The device was in accordance with the specifications at the time of delivery.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.It should be noted that the ifu advices to stop the manipulation if a resistance is felt and advancement by force may result in separation of the dilatation catheter.
|
| |
|
Search Alerts/Recalls
|