BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA LEO 2.5/12; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER Back to Search Results
Model Number 367004
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/15/2019
Type of Reportable Event Malfunction
Event or Problem Description
A pantera leo balloon catheter was chosen for treatment of a severely calcified lesion in a severely tortuous lad.The balloon was inflated up to 20 atm but it did not inflate correctly.It was detected that contrast medium was leaking out the balloon.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed a microscopic leakage and scratches on the balloon surface nearby the damage site.It seems likely that the damage of the balloon surface was caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA LEO 2.5/12
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key9288259
Report Number1028232-2019-04907
Device Sequence Number13855741
Product Code LOX
UDI-Device Identifier07640130415621
UDI-Public07640130415621
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K163660
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model Number367004
Device Catalogue NumberSEE MODEL NO.
Device Lot Number07186015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Initial Date Received by Manufacturer 11/04/2019
Supplement Date Received by Manufacturer12/17/2019
Initial Report FDA Received Date11/06/2019
Supplement Report FDA Received Date12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-