There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler.The incident occurred in (b)(6).Through follow-up communication livanova (b)(4) learned that the device was cleaned regularly per the instruction for use and that the device was placed outside the operation theatre.Through further follow-up communication livanova deutschland learned that the samples are tested randomly once a month and not at the disinfection cycle.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Corrective actions are in progress for this issue.
|