Catalog Number H965700350091
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 10/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.The angiodynamics recent complaint report was reviewed for the stopcock/manifold product family and the failure mode,"package hole/perforated." no adverse trends were identified.The returned sample was visually inspected and the hole in the tyvek portion of the 4 x 8 pouch was confirmed.The damage appears to have been the result of the blue stopcock handle causing the hole.The pouch appeared to have been "severely handled." as part of the receiving process at nipro (the distributor in (b)(6)), all pouched products are removed from their inner boxes and a nipro label (in (b)(6)) is applied to each pouch.Additional handling may occur if product is 100% visually inspected.The pouched product is then reboxed into the inner box by the nipro warehouse employees.The exact root cause of the device pressing against the tyvek in a manner that resulted in a hole cannot be determined.Potential contributing factors include: - handling during transit to nipro warehouse.- handling during nipro labeling/inspection and re-boxing process.All pouches are 100% inspected per angiodynamics procedures during the sealing and final box processes.This is an obvious defect that would have been detected.The directions for use (dfu) packaged with the kits contain the following warning: "contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged." (b)(4).
 
Event or Problem Description
As reported by angiodynamics' distributor in (b)(6), in the distributor's warehouse a hole/tear was found in the tyvek portion of a stopcock pouch, breaching the sterility.The product had not been provided to a hospital and was returned to angiodynamics for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS
Common Device Name
STOPCOCK
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key9289626
Report Number1317056-2019-00147
Device Sequence Number11348244
Product Code DTL
UDI-Device IdentifierH965700350091
UDI-PublicH965700350091
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K782095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2022
Device Catalogue NumberH965700350091
Device Lot Number5503824
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2019
Initial Report FDA Received Date11/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-