| Model Number |
400CE |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer report was received on (b)(6) 2019 ((b)(4)) stating the hemtoherm heater-cooler would not reach maximum temperature.Unit has not been returned for evaluation.Customer reported there was no patient harm.
|
| |
|
Event or Problem Description
|
|
A hemotherm will not reach maximum temperature.Another hemotherm device was used to continue with the procedure without incident.Customer reported there was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was evaluated and determined to have a defective heater.The heater was replaced to correct.
|
| |
|