| Model Number |
N/A |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of asczs0081 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the orange device came off and the needle left the safety base.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of damaged safety mechanism is confirmed but the exact cause could not be determined.One 20 ga x 0.75 in miniloc infusion set was returned for investigation.The device was returned in open packaging with ref: s02020-75 lot: ascz0081.The orange component from the safety mechanism was returned detached from the safety mechanism assembly.Microscopic observation of the safety mechanism assembly did not reveal any apparent mechanical damage.The safety mechanism could not be fully activated.The orange component was able to be reassembled and the safety mechanism was able to be successfully advanced.Based on the condition of the returned sample, possible contributing factors include excessive manipulation of the device and improper assembly.The condition of the device prior to package opening could not be determined; therefore, the complaint is confirmed but the exact cause remains unknown.A lot history review (lhr) of asczs0081 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the orange device came off and the needle left the safety base.No other information was provided.
|
| |
|