• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO L. MED AGGR BLADE (31MMX9MM)S; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO L. MED AGGR BLADE (31MMX9MM)S; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 2296003225
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
 
Event or Problem Description
It was reported that during a surgical procedure, the blade broke at the mount.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
 
Additional Manufacturer Narrative
Additional information: b5: updated procedure details.Device evaluation: d10 & h6: investigation results indicate that contact with metal during use, along with the application of a bending moment could have caused these blades to fail.The ifu of handpieces associated with this device warn the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the cutting accessory.Excessive pressure may bend or fracture the blade and cause tissue damage and/or loss of tactile control." the quality investigation is complete.
 
Event or Problem Description
It was reported that during a commmando surgical procedure, the blade broke at the mount.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L. MED AGGR BLADE (31MMX9MM)S
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9292101
Report Number0001811755-2019-03686
Device Sequence Number2172051
Product Code GFA
UDI-Device Identifier04546540046550
UDI-Public04546540046550
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2021
Device Catalogue Number2296003225
Device Lot Number16284017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2019
Initial Date Received by Manufacturer 10/14/2019
Supplement Date Received by Manufacturer01/09/2020
Initial Report FDA Received Date11/07/2019
Supplement Report FDA Received Date01/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-