| Model Number |
STRIP, RH TMX SUPPLY KIT10CT MG/DL 24/CS |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Codes |
Vomiting (2144); Dizziness (2194)
|
| Date of Event |
10/14/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Meter and test strips were not returned for evaluation.Most likely underlying root cause: mlc-58: user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Note: manufacturer contacted customer in a follow-up call to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 50 and 51mg/dl.The expected fasting blood glucose test result range is 100-150mg/dl.The customer did report symptoms of vomiting and dizziness.Medical attention is reported as a result of the actual blood glucose results and reported symptoms.The product is stored according to specification in the bedroom.During the call a back to back blood test was not performed by the customer.The test strip lot manufacturer¿s expiration date is 01/31/2021 and open vial date is 4 days ago.The meter memory was reviewed for previous test result history.(b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Sections with additional information as of 11-may-2020: h6: updated fda's method, result, and conclusion codes.H10: meter was not returned for evaluation.Test strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested and passed.Corrected sections as of 11-may-2020: h1: type of reportable event corrected from "malfunction" to "serious injury".H3: device was returned to manufacturer for evaluation corrected from "yes" to "no".
|
| |
|
Search Alerts/Recalls
|