Model Number 8FENJ-S
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Extubate (2402)
Date of Event 10/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device's cuff could not enter an air inside.The patient had a replacement of the tube.
 
Additional Manufacturer Narrative
H3 evaluation summary: one sample of tracheostomy tube with evac was received for evaluation.The received components are one fenestrated outer cannula and one obturator.A visual inspection was performed and it was observed all the components are assembled properly at the tube.An inflation/deflation test was performed, with a syringe 15 cc of air was applied through the pvt valve and the inflation/deflation system worked properly, the cuff was inflated and left that way for 24 hours, the valve, inflation line and cuff were submerged into a container filled with water and no leak was observed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key9293370
Report Number9611350-2019-00021
Device Sequence Number2104802
Product Code JOH
UDI-Device Identifier20884521081601
UDI-Public20884521081601
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2023
Device Model Number8FENJ-S
Device Catalogue Number8FENJ-S
Device Lot Number18F0677JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2019
Initial Date Received by Manufacturer 10/23/2019
Supplement Date Received by Manufacturer12/17/2019
Initial Report FDA Received Date11/07/2019
Supplement Report FDA Received Date01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-