| Model Number |
ADVIA 560 HEMATOLOGY SYSTEM |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample barcode was misread on an advia 560 hematology system with an autoloader.Siemens is investigating the issue.
|
| |
|
Event or Problem Description
|
|
A sample barcode was misread on an advia 560 hematology system with an autoloader.Patient sample identification number (sid) (b)(6) was misread as sid (b)(6).There was no other patient sample with the sid (b)(6).Initial results were not reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr 2432235-2019-00410 on 08-nov-2019.Additional information (22-nov-2019): the customer did not provide additional information.The cause of the event could not be established.Potential causes of the event include the customer barcode quality, or that barcode reader programming is needed to accommodate the customer's barcode symbology.The system is performing according to specifications.No further evaluation of this device is required.
|
| |
|