Catalog Number 306499
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the stopper in the bd posiflush¿ sf saline syringe was "thicker" at the end, and the pre-filled solution was "about.4 ml's" under specification.All defects were found during use.The following information was provided by the initial reporter: "the issue is that we have a new version of the posiflush syringe where the rubber stopper at the end is thicker than the previous one, and i was wondering if the dimensions on the new posiflush syringe are different than the old model.Because the ones from the new lot are reading about.4 ml's under.".
 
Additional Manufacturer Narrative
Investigation: two samples were received.One sample is a 306499 and the other one is a 306500.Both came in sealed packaging.Visual inspection was performed finding them both acceptable with no defects.They are different product designs; therefore, the stoppers are different.The 306500 is not anymore produced by our company.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.
 
Event or Problem Description
It was reported that the stopper in the bd posiflush¿ sf saline syringe was "thicker" at the end, and the pre-filled solution was "about.4 ml's" under specification.All defects were found during use.The following information was provided by the initial reporter: "the issue is that we have a new version of the posiflush syringe where the rubber stopper at the end is thicker than the previous one, and i was wondering if the dimensions on the new posiflush syringe are different than the old model.Because the ones from the new lot are reading about.4 ml's under.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SF SALINE SYRINGE
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key9295924
Report Number1911916-2019-01171
Device Sequence Number19337794
Product Code NGT
UDI-Device Identifier00382903064991
UDI-Public00382903064991
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K161552
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number306499
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 10/17/2019
Supplement Date Received by Manufacturer10/17/2019
Initial Report FDA Received Date11/08/2019
Supplement Report FDA Received Date11/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-