| Catalog Number |
306499 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the stopper in the bd posiflush¿ sf saline syringe was "thicker" at the end, and the pre-filled solution was "about.4 ml's" under specification.All defects were found during use.The following information was provided by the initial reporter: "the issue is that we have a new version of the posiflush syringe where the rubber stopper at the end is thicker than the previous one, and i was wondering if the dimensions on the new posiflush syringe are different than the old model.Because the ones from the new lot are reading about.4 ml's under.".
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: two samples were received.One sample is a 306499 and the other one is a 306500.Both came in sealed packaging.Visual inspection was performed finding them both acceptable with no defects.They are different product designs; therefore, the stoppers are different.The 306500 is not anymore produced by our company.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.
|
| |
|
Event or Problem Description
|
|
It was reported that the stopper in the bd posiflush¿ sf saline syringe was "thicker" at the end, and the pre-filled solution was "about.4 ml's" under specification.All defects were found during use.The following information was provided by the initial reporter: "the issue is that we have a new version of the posiflush syringe where the rubber stopper at the end is thicker than the previous one, and i was wondering if the dimensions on the new posiflush syringe are different than the old model.Because the ones from the new lot are reading about.4 ml's under.".
|
| |
|