| Model Number |
7600-0001 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The deflated balloon was the first balloon implanted of the three-balloon system and was implanted for a duration of 179 days.The balloon inflation pressures were recorded as within the acceptable pressure range at implantation and were administered using the ezfill dispenser.Obalon initiated a technical investigation of the product failure including an engineering analysis of the returned balloons.All balloon volumes were within the expected volume specification.The balloons were visually inspected with light microscopy and material fatigue was only observed on the deflated balloon.Scanning electron microscope (sem) images were obtained for the deflated balloon and a breach was identified in the area of the material fatigue.Deflation is a known risk, the frequency of balloon deflations has not exceeded the frequency identified in the labeling for on-label use.
|
| |
|
Event or Problem Description
|
|
During a scheduled endoscopic removal procedure on (b)(6) 2019, two balloons were found inflated and one balloon was found partially deflated floating in the stomach.All balloons were successfully removed by endoscopy and were returned for investigation.The patient did not report a new onset of symptoms to their prescribing physician and had the first balloon placement on (b)(6) 2019, second balloon placement on (b)(6) 2019, and third balloon placement on (b)(6) 2019.
|
| |
|