| Model Number |
700-0015 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The surgeon revised a patient due to lucency on x-ray and thigh pain and replaced the hip stem with a modular, fluted stem.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned part was examined.No x-rays were provided.There was a significant amount of bony tissue adhered to the porous coating of the stem.There were significant score marks on the proximal portion of the stem consistent with osteotome utilization for stem removal.The bony tissue and score marks are suggestive of bony fixation.There were no other indicators or observations that would explain how this device contributed to the reported incident.
|
| |
|