| Catalog Number |
950502089 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Device-Device Incompatibility (2919); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
10/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that after inserting a headed pin through the tc3 4 in 1 block, the pin got struck through the block and distal spacer and could not be removed, both during and after the case.With 20 minutes surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary : examination of the returned device confirmed the reported event.The investigation findings did not indicate that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|